PicoLog 1216 16-kanal USB Data acquisition

Pico Technology
PP544
  • Beskrivelse
  • Mere information
  • Specifikationer
  • 16 unipolar 0 V to 2.5 V analog inputs
  • 12-bit resolution
  • 4 configurable digital outputs
  • 1 PWM control output 
  • Up to 1 MS/s sample rate
  • USB-connected and powered
  • Current-limited sensor excitation output
  • Windows XP (SP2 or later), Vista, Windows 7, 8 and 10 (32- and 64-bit) compatible
  • Includes API and examples for C/C++/C#, VB, LabVIEW VIs
  • Includes ready-to-go data logging software

     

16 channel low cost USB data logger, 12-bit resolution.

(Because the ADC-11/12 was so popular we’ve also added a USB ADC-11/12 compatibility mode which allows you to use your PicoLog 1216 logger as a direct replacement to the USB ADC-11/12).

Designed to meet the needs of a wide range of general-purpose voltage, sensor and transducer logging applications, the PicoLog 1216 Multi-Channel DAQ feature independent software-configurable scaling and control outputs, an optional external terminal board for custom front-end circuitry and a 12-bit input resolution.

A distinguished pedigree
The PicoLog 1216 is the result of a distinguished lineage that goes back to the release of our first multi-channel data logger — the ADC-11/12 — in 1993. The original ADC-11/12, and its successor the USB ADC-11/12, proved to be the perfect choice for users wanting a low-cost way to measure and record multiple signals. The PicoLog 1216 builds on this success to give you the same low-cost data acquisition but with greater power and performance. (Because the ADC-11/12 was so popular we’ve also added a USB ADC-11/12 compatibility mode which allows you to use your PicoLog 1216 logger as a direct replacement to the USB ADC-11/12)

An expandable multi-channel data acquisition system
The powerful PicoLog 1216 has 16 input channels. Need more channels? No problem. Using PicoLog you can connect up to 4 Pico data loggers to one PC — giving you a potential 64 channel PicoLog 1000 Series data acquisition system, or the ability to use your PicoLog 1000 logger with other devices such as the USB TC-08 thermocouple data logger.

Fast and accurate
With 12-bit resolution and multiple sampling modes, a PicoLog 1216 logger will meet your data logging needs. The PicoLog 1216 Series has 3 sampling modes: continuous streaming allows channel voltage readings to be logged continuously at 1 kS/s on any number of channels, while real-time continuous sampling provides averaged, time-accurate readings with automatic measurements available in PicoLog. Block mode captures at the full 1 MS/s sample rate of the logger for the duration of the 8k sample buffer.

Control alarms and other devices
The PicoScope 1216 data loggers include digital outputs — including one with pulse-width modulation control. These outputs can be used to control alarms or other devices, additionally these can also be used to power sensors such as thermistors (outputs can be controlled using the programming API and cannot be controlled using PicoLog).

It’s all you need
Your PicoLog 1216 Series multi-channel data acquisition (DAQ) device includes all you need to start measuring and recording. A full suite of software is included in the price and consists of PicoLog data logging package, the PicoScope oscilloscope package and an SDK for using the PicoLog 1216 within your own software, or third party software such as LabVIEW. Additionally, an optional terminal board with screw terminals lets you easily and quickly connect your sensors to the logger.

Easy to use. Low cost. The PicoLog 1216 is what you need
With great performance and versatility, up to 16 inputs and 4 outputs per unit, a low per-channel cost and easy to use to features, the PicoLog 1216 is the multi-channel data acquisition device that you need.

Inputs
Model PicoLog 1012 PicoLog 1216
Maximum sampling rate
Continuous streaming

Real-time continuous
Block mode#1

1 kS/s per channel in PicoLog, 100 kS/s shared by all channels using the supplied SDK
1 kS/s or greater
1 MS/s single-channel
Buffer memory 8000 samples, shared by all channels
Analog inputs 12#2 16#2
Analog bandwidth (–3 dB) DC to 70 kHz
Input type Single-ended, unipolar
Input voltage range 0 to +2.5 V
Linearity (at 25 °C) 1 LSB
Resolution 10 bits 12 bits
Accuracy 1% 0.5%
Overvoltage protection ±30 V to ground
Input coupling DC
Input impedance 1 MΩ
Outputs
Digital output (D0…D3) 2 4#2
Digital outputs (PWM)
period
duty cycle
None 1
100 µs to 1800 µs
Adjustable from 0% to 100% in 1% steps
Digital outputs (all)
logic low voltage
logic high voltage
current limit

100 mV (typical)
3.3 V
1 kΩ resistors in series with outputs
Power output for sensors 2.5 V @ 10 mA, current-limited
PC requirements
Minimum Processor: 1 GHz.
Memory: 512 MB.
Free disk space: 32-bit: 600 MB, 64–bit: 1.5 GB.
Operating system: 32- or 64-bit edition of Microsoft Windows XP (SP3), Vista, Windows 7 or Windows 8 (not Windows RT).
Beta software: Linux Debian, OpenSUSE and Fedora derived distributions, OS X (Mac).
Ports: USB 1.1 port.
Recommended Processor: 1 GHz.
Memory: 512 MB.
Free disk space: 32-bit: 850 MB, 64–bit: 2 GB.
Operating system: 32- or 64-bit edition of Microsoft Windows 7 or Windows 8 (not Windows RT).
Beta software: Linux Debian, OpenSUSE and Fedora derived distributions, Mac OS X.
Ports: USB 1.1 or higher.
Environmental
For quoted accuracy 20 °C to 50 °C
General operation 0 °C to 70 °C
Relative humidity 5% to 80% RH
Physical properties
Dimensions 45 x 100 x 140 mm (1.77 x 3.94 x 5.51 in)
Weight <200 g (7.05 oz)
Software
PicoLog for Windows PicoLog data acquisition software can collect up to 1 million samples. Features include:
Multiple views — view data as a graph, spreadsheet or text
Parameter scaling — convert raw data into standard engineering units
Math functions — use mathematical equations to calculate additional parameters
Alarm limits — program an alert if a parameter goes out of a specified range
IP networking — transfer measurements via a LAN or over the Internet
Operating systems supported: Microsoft Windows XP (SP3) to Windows 10. Not Windows RT.
PicoLog languages Full support for: English, Français, Deutsch
Menus and dialogs only for: Italiano, Español, Svenska
PicoScope 6 for Windows PicoScope 6 is your complete test and measurement lab in one application. Features include:
Capture modes: oscilloscope, spectrum and persistence.
Channel maths: calculate the sum, difference, product, inverse or create your own custom function using standard arithmetic, exponential and trigonometric functions.
Mask limiting testing: pass/fail, failure count, total count.
Serial Decoding: decode data from a serial bus such as I²C.
Automated measurements
Scope mode: AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width, low pulse width, maximum, minimum, peak–to–peak, rise time and rising rate.
Spectrum mode: frequency at peak, amplitude at peak, average amplitude at peak, total power, total harmonic distortion (THD % and THD dB), total harmonic distortion plus noise (THD+N), spurious-free dynamic range (SFDR), signal+noise+distortion to signal+noise ratio (SINAD), signal to noise ratio (SNR) and intermodulation distortion (IMD).
Export data formats: comma separated values (CSV), tab delimited (TXT), Windows bitmap (BMP), graphics interchange format (GIF), portable network graphics (PNG), MATLAB 4 format (MAT).
Operating systems supported: Microsoft Windows 7, 8 and 10* (not Windows RT).
PicoScope languages Chinese (simplified), Chinese (traditional), Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Spanish, Swedish, Turkish
Software Development Kit Includes example code.
Optional free software PicoScope 6 Beta for Linux and OS X. Drivers for Linux and OS X.

* PicoScope R6.11 and earlier and the SDKs are compatible with Microsoft Windows XP (SP3) and Vista SP2 as well as the versions mentioned above.

Documentation languages
Documentation
User’s guide
Programmer’s guide
Terminal board user’s guide

English, Français, Deutsch, Español
English
English
General
Additional hardware (supplied) USB 2.0 cable, user manuals, software CD-ROM
PC interface USB 2.0
I/O connector 25-way D female
Power requirements Powered from USB port
Ground fault current protection 0.9 A thermal self-resetting fuse
Compliance European EMC and LVD standards
FCC Rules Part 15 Class A
RoHS compliant

#1Block mode is supported by PicoScope and the PicoLog 1000 Series SDK, but not by PicoLog.

#2Except in USB ADC-11 compatibility mode.